Automatic Calibration/Adjustment with isoCAL
Fully automatic calibration/adjustment ensures that the balance is automatically calibrated and adjusted when predefined time intervals or temperature values are exceeded.
This function varies in different models:
Entris / Quintix / Secura: isoCAL is automatically carried out based on time or temperature, and each time the balance is leveled. This function can be switched off via the system settings
For information: In Secura models, every adjustment process, including the identified deviations, is documented and saved on the balance. The saved reports can be viewed via the system settings and printed out on a PC or Sartorius lab printer.
⊳ If the ambient temperature has changed since the last calibration/adjustment or the preset time interval has been exceeded, the red isoCAL button appears on the display.
⊳ If the isoCAL function is set to Automatic, the calibration/adjustment process starts automatically.
Δ If the balance has not been adjusted, the prompt will be shown again 5 minutes after the “isoCAL” status message is acknowledged.
⊳ If the isoCAL function is set to Info, manual start, the function must be started manually.
⏵ To start the calibration/adjustment function manually, select the isoCAL button (Entris, Secura, Quintix).
⊳ Calibration starts automatically.
⊳ A report appears on the display (Secura and Quintix only).
The report indicates the deviations identified during isoCAL calibration.
The report also gives the result of the adjustment process.
⏵ Close the report window: Select X.
⊳ The balance is now calibrated and adjusted